  • To avoid exposure of the resist by ambient light, we recommend yellow light illumination.
  • If an optical table was ordered from us, compressed air with a pressure between 5bar and 8bar and a connector for 6mm outer diameter tubing has to be supplied.

For the working conditions of the system please refer to the following list:

 

Electrical Properties

mains voltage                           AC 230 V (+/- 10 %); 
                                             US: AC 120 V (+/-10 %)
mains frequency                        50 - 60 Hz
maximum power consumption       2000 VA
mains supply overvoltage            Category II
grounding equipment conductor    required
electric safety                          according to DIN EN 61010-1
                                             (IEC 61010-1)

 

Ambient Conditions

operating conditions                    indoors
operating temperature                 22° C (+/- 4° C)
temperature stability                   < 1° C overall fluctuation
storage temperature                     20° C (+/-20° C)
relative humidity                          maximum 60 % up to 30° C
operational altitude                      ≤2000m
active vibration isolation               required

 

Weights and Measures

total weight                               100 kg
dimensions (table)                       90 cm x 90 cm x 60 cm
dimensions (rack)                        56 cm x 60 cm x 64 cm 

 